Adverse events of two vessel-sealing devices in robotic urologic surgery · FDA MAUDE database analysis · Maimonides Medical Center · MSRE, summer 2024
FDA MAUDE database analysis · Maimonides Medical Center · MSRE, summer 2024

Adverse events of two vessel-sealing devices in robotic urologic surgery

Henry M. Blair · Christopher Edwards · Ariel Schulman, MD · Jacob Khurgin, MD

Robotic vessel sealers cut and seal blood vessels in nearly every major urologic operation. On March 3rd, 2022 the FDA issued a field safety notice about two of them — SynchroSeal and Vessel Sealer Extend — warning that seals could fail when excessive tissue is caught in the jaws. A safety notice says a failure mode is possible; the FDA's adverse-event database says what has actually been reported.

67
urologic adverse-event reports reviewed
p = 0.025
severity difference between devices
22% / 19%
sealing / cutting issues, SynchroSeal
Status. Conducted through the Medical Student Research Experience at Maimonides in summer 2024 and presented at the Society of Urologic Robotic Surgeons Focal Therapy Conference, 2024, and at regional surgical meetings.

The question

What adverse events are reported for the SynchroSeal and Vessel Sealer Extend in urologic surgery, do the two devices differ in the severity of what is reported, and do the reports match the failure mode the FDA warned about?

How it is being done

What it shows so far

Why the two-axis classification mattered. Ranked by raw report count, one device looks worse. Ranked by severity of events actually attributable to the device, the ordering reverses. Reporting volume is not a safety signal.

Materials

What comes next

The queries were restricted to urologic reports for time. Extending to all surgical specialties would add the statistical power this analysis lacks and test whether the severity difference holds outside urology.